• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HST III SYSTEM (3.8MM); CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HST III SYSTEM (3.8MM); CLAMP, VASCULAR Back to Search Results
Model Number HST III SYSTEM (3.8MM)
Device Problems Fitting Problem (2183); Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2022
Event Type  malfunction  
Event Description
Related to (b)(6).The hospital reported that the hst iii system (3.8mm) occluder does not release.The diameter of the blocking umbrella is larger than the diameter of the delivery tube, so it cannot be released.The blocking umbrella is in the delivery pipe.No surgical delay.Replace the same equipment to complete the surgery.No harm to the patient.
 
Manufacturer Narrative
(b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Manufacturer Narrative
Trackwise # (b)(4).The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HST III SYSTEM (3.8MM)
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key14348147
MDR Text Key291315898
Report Number2242352-2022-00401
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00607567700314
UDI-Public00607567700314
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/26/2022
Device Model NumberHST III SYSTEM (3.8MM)
Device Catalogue NumberC-HSK-3038
Device Lot Number25159821
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/27/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/26/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
Patient Weight68 KG
-
-