• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE¿ EXTENSION SET, BIFUSE 2 SMARTSITE¿ NEEDLE-FREE CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE¿ EXTENSION SET, BIFUSE 2 SMARTSITE¿ NEEDLE-FREE CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 20159E
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is unknown; awareness date has been used for this field.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd alaris smartsite¿ extension set, bifuse 2 smartsite¿ needle-free connector the needle was clogged.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer that there are issues flushing meds through the extension sets.They have issues flushing meds through them.
 
Additional Manufacturer Narrative
H6: investigation summary: two samples were returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The set was attached to a cut off bd syringe to see if a fluid path can be observed while the piston is in the activated position.A fluid path was observed while the piston was in the activated position.The smartsite was then connected to a 10 ml bd syringe filled with blue dye water and the set was attempted to be flushed.The set was successfully flushed.The customer complaint that there are issues flushing meds through the extension sets could not be replicated.The root cause could not be determined because the issue could not be replicated.A device history record review for model 20159e lot number 22029775 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 22feb2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A trend for this occlusion issue has been identified for this product line.A capa (corrective action preventative action) 1998036 has been initiated, and a team has been assembled in order to investigate the issue.
 
Event or Problem Description
It was reported while using bd alaris smartsite¿ extension set, bifuse 2 smartsite¿ needle-free connector the needle was clogged.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer that there are issues flushing meds through the extension sets.They have issues flushing meds through them.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE¿ EXTENSION SET, BIFUSE 2 SMARTSITE¿ NEEDLE-FREE CONNECTOR
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14349170
Report Number9616066-2022-00569
Device Sequence Number20063974
Product Code FPA
UDI-Device Identifier37613203012288
UDI-Public37613203012288
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/17/2025
Device Model Number20159E
Device Catalogue Number20159E
Device Lot Number22029775
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date05/10/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-