• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA LLC PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN); PUMP, BREAST, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA LLC PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN); PUMP, BREAST, POWERED Back to Search Results
Model Number 101041361
Health Effect - Clinical Code Bacterial Infection (1735)
Date of Event 04/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer service conducted troubleshooting including; verifying there was no milk backup; disassembling, inspecting, cleaning and reassembling kit and tubing set; and verifying correct breast shield size.The customer was sent a replacement pump and return of her original pump was requested for testing/evaluation.In follow up with a complaint handler on 04/21/2022, the customer indicated that she developed mastitis on (b)(6) 2022 and was prescribed an antibiotic.On 04/27/22, the customer responded to the complaint handler and she indicated that the replacement pump was working without issue and her mastitis was resolved.The device was returned with the customer's parts and accessories and was evaluated on 05/04/2022.The device was tested with the customer tubing and lab kit(connectors and membranes).There was human milk residue found on the pump housing, pump aggregate (pump motor and pump assembly), chassis, and manifold.The customer's report of low suction could be confirmed.Based on the results of our internal investigation (reference number (b)(4)),it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is 0.008% for the period of january 2013 to august 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history of mastitis.Riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.However in this case, the mother¿s mastitis required prompt medical attention for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Event or Problem Description
On (b)(6) 2022, the customer alleged to medela llc that she had no suction with her pump in style max flow breast pump.She additionally alleged she had mastitis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN)
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer Contact
dave kurudza
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key14349279
Report Number1419937-2022-00046
Device Sequence Number14491926
Product Code HGX
UDI-Device Identifier00020451401454
UDI-Public020451401454
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K200508
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101041361
Device Catalogue Number101041361
Device Lot Number0000038430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2022
Distributor Facility Aware Date04/21/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2022
Initial Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-