|
According to the reporter, during a laparoscopic gastric gastrointestinal stromal tumors procedure, while transecting the gastric, and when the surgeon tried to do a double stapling on top of the staple line, the device failed to fire.Even after changing to another reload, the issue still occurred.The surgeon did hand suturing to complete the case and resolve the issue.There was no patient injury.
|
|
D10 concomitant product: egiaustnd, egiaustnd endogia ultra univ std stap (lot#p1j0505).H3 evaluation summary: medtronic conducted an investigation based upon all information received.Two devices were available for evaluation.Visual inspection noted the device was partially fired and the product's jaws would not close completely due to a deformed clamping mechanism.It was reported that the device did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The issue can occur in any of the following circumstances: 1.Firing over tissue that is beyond the recommended thickness range.2.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|