The meter and strips were requested for investigation.Replacement product was sent.The product is not expected for return or has not been received at this time.If the product is returned in the future, a follow up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.".
|
|
We received an allegation of questionable inr results for one patient tested on a coaguchek xs meter serial number up1071193 compared to an unknown laboratory methodology.On (b)(6) 2022 and at 1:35 p.M., the patient's meter result was 7.2 inr.At 2:13 p.M., the patient's laboratory result was 4.7 inr.On (b)(6) 2022 and at 9:56 a.M., the patient's meter result was 5.8 inr.At 12:20 p.M., the patient's laboratory result was 4.4 inr.The patient's therapeutic range was 2.0-3.0 inr.The patient's testing frequency is every two weeks.
|