| Model Number |
SU-XXXXX |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.At the submission of the case, the customer has provided picture of the atrasump cannulae complained for the disconnection.Complained cannula that detached during use is available and have been requested for investigation.Unit has not been received at livanova facilities.If any additional information pertinent to the reported event is obtained, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova usa received a report of atrasump tip disconnection during its placement in the heart.There is no report of any patient injury.Additionally, it was reported that in another atrasump cannula belonging to the same item code disconnected from the cannula body during manipulation of the cannula to position in the heart before starting the procedure.The second unit has been disposed by the customer.
|
| |
|
Event or Problem Description
|
|
See intial report.
|
| |
|
Additional Manufacturer Narrative
|
|
Livanova received a report stating that, during the procedure, the tip of the su-29602 suction sump cannula came off.No patient was affected.The dhr review highlighted that the complained lot was released conforming to specifications.Furthermore, it was reported that a second cannula belonging to the same lot had the same issue that occurred before starting the procedure.Thus, in total two units involved.The bloodstained unit, the only one available, was requested for further investigation but was never shipped to livanova arvada.Therefore, the evaluation of the event is based on previous similar events.Review of the livanova complaint database identified in total 4 similar events related to this type of cannula.Inspection of the devices of previous events suggested that the seal at the disconnected part (site between the connector and the tubing) has been done with a thin layer of adhesive.Dimensional analysis of the critical dimensions for this connection were found to be in specification.Based on livanova investigation, it cannot be ruled out that the root cause of the tip detachment is related to a manufacturing deviation during the assembly phase.To prevent reoccurrence, a capa project was launched.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See intial report.
|
| |
|
Additional Manufacturer Narrative
|
|
In the frame of the capa project, additional investigation of the bond between the connector and tubing has indicated a strength limit related to the adhesive bond between these components due to the design of the connector.To increase the tip to tube pull strength, design changes of the connector are under investigation.The risk remains in the acceptable region.Livanova will keep monitoring the market.Investigation conclusion has been updated accordingly.
|
| |
|