| Model Number |
SSM-140 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device related to the reported event of capsular contracture was received on july 19, 2022 with lot number 3493937.Visual analysis of the returned device identified flat creases.A weight test of the device was verified and the device was within specification.Cloudy after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: -no issues found related with the manufacturing.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|