| Model Number |
71676627 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during internal fixation, an intertan 1.5 13x42 130d lt was opened, and it appeared normal.When attached to the targeter, it would only attach in the ¿reverse¿ as a right nail would.Further, inspection showed a lime green nail, clearly marked left that appears to actually be a right nail incorrectly colored and labeled.Replacement/alternate had to be retrieved.Surgery was resumed after a non-significant delay the procedure was finished with a smith and nephew back up device.The patient was not harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
H7: if remedial action initiated, check type and h9: if action reported to fda under 21 usc 360i(f), list correction/removal reporting number.Section h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The orientation of the device was determined to be incorrect according to it labeling.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.According to the drawing print, the device for the left side should have ¿left¿ laser etch.At this time, we do have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that this event was previously identified and addressed through a training program, a workshop and the review of existing line clearance instructions.A contribution of the device to the reported event could be corroborated as the device was not in the correct orientation.A potential probable cause could be but is not limited to a mix-up of two orders prior to colour anodizing.Based on this investigation, the need for corrective action is indicated.As a result of the investigation, several process improvements were implemented related to process monitoring and in process inspection.This event was addressed through a market removal of the product.Containment and remedial actions were performed for the impacted finished product.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|