SMITHS MEDICAL ASD, INC. BIVONA; TRACHEOSTOMY TUBE AND TUBE CUFF Back to Search Results
Model Number FT21HN40NGE133N
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No device was returned for investigation.If the device is received the investigation will be reopened.
 
Event or Problem Description
It was reported that the device leaks from the inside at the adapter end.No additional information.
 
Additional Manufacturer Narrative
Additional information received.Updated section b3.
 
Additional Manufacturer Narrative
Other text: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found did not reveal any anomalies.Functional testing found a slow, inside leak was detected at the lumen by the connector.The investigation determined that there was insufficient adhesive inserted into the lumen above where the airway line is secured to the lumen.Root cause was determined to be manufacturing.To heighten awareness, a quality alert was issued to the production floor for custom lab operator review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TRACHEOSTOMY TUBE AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14355854
Report Number3012307300-2022-08504
Device Sequence Number20063912
Product Code JOH
UDI-Device Identifier15019517279122
UDI-Public15019517279122
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberFT21HN40NGE133N
Device Catalogue NumberFT21HN40NGE133N
Device Lot NumberGA02654
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer05/20/2022
07/05/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date06/14/2022
08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age2 YR
Patient SexFemale
-
-