BECTON DICKINSON MEDICAL (SINGAPORE) BD ECLIPSE¿ NEEDLE WITH DETACHABLE BD LUER-LOK¿ SYRINGE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305780
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple device names/types and 510k numbers reported to be involved.Additional information below: medical device type: fmf.Common device name: piston syringe.Pma / 510(k)#: k941562.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the stopper was found misaligned on the bd eclipse¿ needle with detachable bd luer-lok¿ syringe plunger before use.The following information was provided by the initial reporter: "distorted stopper incident occur : before use".
 
Additional Manufacturer Narrative
H6: investigation summary: one photo was received by our quality team for evaluation.From the photo, the team observed that the plunger is withdrawn to the approximate scale marking 0.62ml and there are signs of distortion at the top stopper edge.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.The 1ml syringe part is manufactured at bd canaan and the quality team there has conducted a supplemental investigation on this nonconformance.A potential root cause for the jammed stopper defect is associated with the assembly process.H3 other text : see h10.
 
Event or Problem Description
It was reported that the stopper was found misaligned on the bd eclipse¿ needle with detachable bd luer-lok¿ syringe plunger before use.The following information was provided by the initial reporter: "distorted stopper incident occur : before use.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d.10.Device available for eval: yes d.10.Returned to manufacturer on: 20-jul-2022 h.6.Investigation summary: one photo and one sample were received by our quality team for evaluation.From the photo, the team observed that the plunger is withdrawn to the approximate scale marking 0.62ml and there are signs of distortion at the top stopper edge.The sample was subjected to visual inspection and the top edge of the stopper is observed to be distorted downwards inside the syringe barrel.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.The 1ml syringe part is manufactured at bd canaan and the quality team there has conducted a supplemental investigation on this nonconformance.A potential root cause for the jammed stopper defect is associated with the assembly process.H3 other text : see h.10.
 
Event or Problem Description
It was reported that the stopper was found misaligned on the bd eclipse¿ needle with detachable bd luer-lok¿ syringe plunger before use.The following information was provided by the initial reporter: "distorted stopper incident occur : before use".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ECLIPSE¿ NEEDLE WITH DETACHABLE BD LUER-LOK¿ SYRINGE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14356512
Report Number8041187-2022-00247
Device Sequence Number3987966
Product Code FMI
UDI-Device Identifier30382903057802
UDI-Public30382903057802
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K161170
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2026
Device Model Number305780
Device Catalogue Number305780
Device Lot Number1237349
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer05/26/2022
08/15/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date06/24/2022
09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-