Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2018, the patient underwent an unknown surgery for proximal femoral nail antirotation (pfna).The surgery was completed successfully without any surgical delay.On (b)(6) 2022, it was confirmed that cut-out occurred.The revision surgery was done on (b)(6) 2022.The pfna was removed and replaced with an artificial head.The surgeon commented that femoral neck fracture was the cause of the cut-out.It is unknown when the femoral neck fracture occurred.This is report 3 of 4 for complaint (b)(4).This report is for pfna end cap.
 
Manufacturer Narrative
Additional narrative: event date: event is unknown.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Device is not distributed in the united states, but is similar to device marketed in the usa.No code is for revision surgery and injury the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: part #: 473.170s, lot #: l568150, manufacturing site: werk bettlach, supplier: n/a, release to warehouse date: 25 sep 2017, expiration date: n/a.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the pfna-ii end cap extens.0 tan.The provided radiographic image was reviewed and no defects or damages were observed with the end cap.A dimensional inspection was not performed for the pfna-ii end cap extens.0 tan due to the complex geometry of the device.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the pfna-ii end cap extens.0 tan was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFNA-II END CAP EXTENS. 0 TAN
Type of Device
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14356626
MDR Text Key291384672
Report Number8030965-2022-03124
Device Sequence Number1
Product Code HSB
UDI-Device Identifier07611819837652
UDI-Public(01)07611819837652
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number473.170S
Device Lot NumberL568150
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2017
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - NAIL HEAD ELEMENTS: PFNA BLADE; UNK - NAILS: PFNA
Patient Outcome(s) Required Intervention;
Patient Age87 YR
Patient SexFemale
-
-