EXACTECH, INC. PS TIBIAL INSERTS SZ 4, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
PS TIBIAL INSERTS SZ 4, 13MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: fluted stem extension 120l x 14 mm (cat# 204-34-12 / serial# (b)(4)).Tibial stem ext.Screw (cat# 204-70-00 / serial# (b)(4)).Logic pts, size 4, 8mm (cat# 02-012-42-4008 / serial# (b)(4)).Lgc tibial fit tray cem sz 4f / 3t (cat# 02-012-45-4030 / serial3 (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately fifteen years post initial left tka, the (b)(6) male patient had a full revision going to a constrained knee.Patient's revision was due to instability of femoral component.There was poly wear and metallosis due to the 8mm tibial spacer and osteolysis was also present.Patient was revised to a size 4 left constrained femur with splined stems in the femur and tibia as well as metaphyseal cones and a 25 mm constrained liner.There was no device breakage or surgical delay/prolongation.Patient was last known to be in stable condition following the event.X-rays and photos received.The devices will be returned.As of note: "patient was revised in 2014 with loosening of the tibia and the size 4/13mm posters stabilized tibial insert reference number 204¿24¿13 with serial number (b)(4).Is a non-conformed insert that it is on the list of polys".
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the revision reported was likely the result of patient-related conditions including the fall reported, malalignment, instability, third body wear, or any combination of these possibilities, which led to prosthesis wear, osteolysis, and metallosis.Additionally, a contributing factor to the wear and delamination may have been the result of being packaged in a non-conforming vacuum bag for more than five years.The most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
|
| |
|