Other, other text: no lot number was provided; therefore, device history record review could not be performed.Two (2) product samples were received for evaluation.Visual inspection was performed.Visual inspection found the samples were stressed, as the part exhibited white marks around the components.The root cause of the reported issue was found to be that it could not be attributable to the manufacturing site; the most likely cause of the condition observed on returned samples indicate a possible mishandling during use, as force is required to break the devices.
|