| Model Number |
3L81-23 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.Patient identifier complete sid: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely increased creatinine result generated on the architect c8000 analyzer on a patient.Results provided: 15apr2022 sid (b)(4) (normal range: 0.6-1.3 mg/dl), lot# 09277un21.Serum sample: sn# (b)(4) = 2.9 / 2.9 mg/dl, sn# (b)(4) = 2.9 / 2.9 mg/dl, sn# (b)(4) = 2.8 mg/dl, sn# (b)(4) = 2.8 mg/dl.Previous result for this patient = 0.9 mg/dl.Mayo labs (roche platform) = 0.9 mg/dl no impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets determined that there was no other complaint activity for creatinine (ln 3l81-23).Trending review determined no trends for falsely elevated results for the product.Return testing was not completed as returns were not available.A review of the product labeling concluded that the issue is sufficiently addressed.Use error may have contributed to the customer¿s issue as a plasma sample, drawn at the same time did not show the higher-than-expected results as did the serum sample.Based on the investigation, no systemic issue or deficiency for the creatinine was identified.
|
| |
|