• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MEDICAL INCORPORATED NB3 LED PHOTOTHERAPY, 115V, US, LIGHT ASSEMBLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MEDICAL INCORPORATED NB3 LED PHOTOTHERAPY, 115V, US, LIGHT ASSEMBLY Back to Search Results
Model Number 001103
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
Nb3 led phototherapy, 115v, us, light assembly - missing/bent screw and bent roll stand arm.No injuries reported.
 
Additional Manufacturer Narrative
Initial report ref to natus complaint#- (b)(4).Missing/bent screw and bent roll stand arm - neoblue unit.Related to ae- (b)(4) the customer noted they have another neoblue unit that was in use that has a bent arm and a missing screw on the device.Customer stated no harm came from this 2nd device, but they noted: " that the light bank is supposed to pivot, but on many of the neoblues the pivot feature is very rigid (causing the staff to push up and down on it very hard) and that also some of the pivot features are very lax, causing the lights to swing down when not supported by an incubator top or other device.This may contribute to strain on the screws and bolts that secure the light/arm to the stand.Natus investigation team will wait for the customer to return the product for evaluation to verify the alleged customer complaint.There are no capa's related to this issue and this complaint does not identify a deficiency in the product design and therefore a capa is not required.Per qms-004442 complaint histories are reviewed routinely per quality system requirements and any complaint trends are assessed and documented as part of these reviews.No complaint trends have been identified.Service repair investigations will be conducted during the repair process and trend data will be reviewed per qms-004442.Install date: 14-mar-2018.
 
Additional Manufacturer Narrative
Follow up report 001ref to natus complaint# (b)(4).Related to complaint# 10154485 (mdr 9612330-2022-00020).Pictures were provided.Engineering reviewed pictures in the complaint and compared to the roll stands in house.Technical service sent the customer rework instructions for nb3 stand and requested information on who may have assembled the nb3 units.Technical service advised to use instructions to ensure the units they currently have are assembled correctly per the instructions.Customer responded they were unable to tell technical service who initially assembled the units.The affected parts were returned for investigation and the results were as follows: returned neoblue overhead had roll stand arm still attached to it.Arm wasn't attached to the roll stand anymore.Some parts (which include bolts that secure arm to roll stand) were returned in a bag.Threading on roll stand holes and threading on bolts looked fine (not stripped).Bolt that is in "mushroom cap" is mispositioned (tipped).Feet on the neoblue overhead have torn off or were missing.No other visible damage.Root cause/probable root cause: assessment: bolt that's in "mushroom cap" must have been removed by staff at hospital (it's tool removable only, and should never be removed).Recommended actions: customer provided with rework instructions by technical service and instructed to ensure all units currently on site are assembled accordingly.Faillure confirmed:yes.Closure rationale: complaint verified, isolated incident and monitor for future occurrence.Low safety risk of harm, individual complaint related to issue stated.Complaint will be included in trending data for further review and investigation if needed.
 
Event or Problem Description
Nb3 led phototherapy, 115v, us, light assembly - missing/bent screw and bent roll stand arm.No injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NB3 LED PHOTOTHERAPY, 115V, US, LIGHT ASSEMBLY
Common Device Name
NB3 LED PHOTOTHERAPY, 115V, US, LIGHT ASSEMBLY
Manufacturer (Section D)
NATUS MEDICAL INCORPORATED
dba excel-tech ltd. (xltek)
2568 bristol circle
oakville, ontario l6h 5s1,
CA 
Manufacturer (Section G)
NATUS MANUFACTURING LIMITED
ida business park
gort
galway h91pd92,
EI  
Manufacturer Contact
walsh
dba excel-tech ltd. (xltek)
2568 bristol circle
oakville, ontario l6h 5s1, 
CA  
MDR Report Key14358154
Report Number9612330-2022-00025
Device Sequence Number17179515
Product Code LBI
UDI-Device Identifier00382830011594
UDI-Public00382830011594
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number001103
Device Catalogue Number001103
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer04/12/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-