TELEFLEX MEDICAL SDN. BHD. FOLEY CATHETERS; CATHETER, RETENTION TYPE, BALL Back to Search Results
Model Number IPN050125
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).The device has not been returned for investigation.
 
Event or Problem Description
The probe breaks when installed on the patient.Sometimes after a few hours and sometimes after a few days.They broke on sleeping, sedated little babies who did not move.The patients who had the catheters in place were all small, intubated patients under 8kg.The catheters were all size 6 french.The last patient was a cardiac surgery patient who was intubated, sedated and critical.Less than 3kg.In all cases the parts were all removed without obvious complications.New catheters were installed as was necessary to the patient's condition with extra surveillance by the nurse.I cannot recall if there were any urinary infection prophylactic antibiotics given to the patients on related to the breaks.
 
Event or Problem Description
The probe breaks when installed on the patient.Sometimes after a few hours and sometimes after a few days.They broke on sleeping, sedated little babies who did not move.The patients who had the catheters in place were all small, intubated patients under 8kg.The catheters were all size 6 french.The last patient was a cardiac surgery patient who was intubated, sedated and critical.Less than 3kg.In all cases the parts were all removed without obvious complications.New catheters were installed as was necessary to the patient's condition with extra surveillance by the nurse.I cannot recall if there were any urinary infection prophylactic antibiotics given to the patients on related to the breaks.
 
Additional Manufacturer Narrative
Qn#(b)(4).The device lot number was not provided; therefore a dhr review could not be conducted.There was no complaint sample returned for investigation, only photo available.Therefore, no physical assessment could be conducted.In the absence of any actual or representative sample for investigation, we could not further investigate to identify the actual root cause of this reported failure.Therefore, this complaint could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLEY CATHETERS
Common Device Name
CATHETER, RETENTION TYPE, BALL
Manufacturer (Section D)
TELEFLEX MEDICAL SDN. BHD.
perak, west malaysia
Manufacturer (Section G)
TELEFLEX MEDICAL SDN. BHD.
lot no : pt2577 jalen perusahaan
4 kamunting industrial estate
perak, west malaysia 34600
MY   34600
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key14358414
Report Number8040412-2022-00134
Device Sequence Number4155170
Product Code EZL
UDI-Device Identifier04026704307968
UDI-Public04026704307968
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K972184
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN050125
Device Catalogue Number170003060
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-