| Model Number |
U411 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer visually checked the urine sample because the blood for the first sample seemed to be positive.The customer performed a reset and the sample ids were correct.All hardware and software parts were checked and no issues were found.The investigation found the first sample in the customer's list did not have a sample id, and there was a gap in the list with the sample sequence numbers.It is unknown if there were any error messages.Based on the provided information, the investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a software issue for the cobas u 411 urine analyzer.The alleged issue was a mismatch of sample ids.A sample id was displayed as "supervisor" for the first sample instead of the correct sample id.The second sample result had the first sample's sample id.The result, displayed for the second sample, belonged to the first patient sample.The remaining samples on the list were also mismatched.The customer discovered the issue by comparing the results with a new urisys 1100.The results were not reported outside of the laboratory.The software version is 3.3.2.0007.
|
| |
|