| Model Number |
406702 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the procedure, it was noted that there was blood flow back in the sheath and the flush port was warped.The port was clamped with a pair of hemostats to stop the back flow and the case was completed with no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One 7f fast-cath introducer sheath was received for evaluation.The extension tubing had detached from the sideport of the hemostasis body due to an insufficient amount of solvent on the extension tubing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
|
| |
|