Manufacturer's investigation conclusion: review of the submitted log files confirmed multiple pulsatility index (pi) events.The report that the pi events were due to pump position could not be confirmed through this evaluation.It was reported that the patient looked good and was stable per their baseline.The controller event log file contained information from (b)(6) 2022.Multiple pi events were captured throughout the file.No other notable events or alarms were captured.The pump appeared to function as intended at the set speed.Multiple attempts were made to obtain additional information from the customer regarding the event; however, no additional information was provided.The patient remains ongoing on heartmate 3 left ventricular assist system (lvas), serial number (b)(6), with no further issues reported at this time.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specification.The heartmate 3 left ventricular assist system (lvas) instructions for use (ifu), rev.C, is currently available.Section 1 of the ifu, ¿introduction¿, provides an explanation of all pump parameters, including pulsatility index (pi).Section 4, ¿system monitor¿, explains that pi events are assumed by the system during cases when there are sudden and substantial changes in the pulsatility index.These events are also referred to as pi events, and may be initiated for reasons other than true pi events.Some reasons include sudden changes in a patient¿s volume status, arrhythmias, sudden changes in power, and sudden changes in pump speed.If the system detects a pi event, the pump speed automatically drops to the low speed limit and slowly ramps back up at a rate of 100 rpm per second to the fixed speed setpoint.Section 6, "patient care and management¿, explains that pump performance is sensitive to changes in systemic vascular resistance and left ventricular filling.Section 5 of the ifu, "surgical procedures," explains how to insert the pump in the ventricle and instructs the user to take care to avoid orienting the inlet towards the interventricular septum as pump function will be compromised in the presence of inlet obstruction.The steps to adjust the orientation of the pump are also outlined in this section.No further information was provided.The manufacturer is closing the file on this event.
|