Model Number 1804350-18
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo, 90% stenosed, heavily calcified and heavily tortuous vessel in the left circumflex (lcx).The 3.5x18mm xience skypoint stent delivery system (sds) met resistance with the lesion during advancement and failed to cross the lesion due to the anatomy.The sds was removed independently.Resistance was met with the lesion during removal.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the device interacted with the 90% stenosed, heavily calcified and heavily tortuous vessel causing the reported failure to advance.Continued resistance with the difficult anatomy was experienced during removal causing the reported difficulty to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14360033
Report Number2024168-2022-05151
Device Sequence Number17357718
Product Code NIQ
UDI-Device Identifier08717648233401
UDI-Public08717648233401
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2024
Device Model Number1804350-18
Device Catalogue Number1804350-18
Device Lot Number111014A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer05/13/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-