|
WRIGHTS LANE SYNTHES USA PRODUCTS LLC PROTECTION SLEEVE FOR FNS INSRT INSTS; CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Model Number |
03.168.013 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in the usa as follows: it was reported that on an unknown date, the parts of protection sleeve tip was broke/missing.There was no patient involvement.This complaint involves one (1) device protection sleeve for fns instr inst's.This is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e3.Product: 03.168.013.Lot : 190453-401.Release to warehouse date: june 05, 2020.Manufacturing site: werk selzach.Suppliers: leitner ag.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the protection sleeve for fns insert insts was found the device broken on the tip.The broken fragment was not returned.No other defect was found.A dimensional inspection for the protection sleeve for fns insert insts was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the protection sleeve for fns insert insts would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: the drawings reflecting the current and manufactured revisions were reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|