Manufacturer contacted the physician to obtain additional information related to this event.Upon analysis of event, it has been concluded that the thermal effect is highly unlikely to be a result of fluid leak at the cervix.This conclusion was based on several factors: 1.Fluid loss was never observed by physician.2.Fluid loss was never reported by the respective feature of the genesys system 3.If fluid loss would be the cause, the thermal effect would involve the vaginal mucosa first and would likely affect the posterior wall of the vagina.Based on the above and considering the lateral labial location of the above-named effect, it is more likely that thermal energy affected the tissue secondary to direct contact between the sheath of the device and the skin.While the physician reportedly did not remove the speculum during the ablation cycle, it is possible that some of the tissue inadvertently established temporary contact with the sheath, resulting in unintended thermal effect.This scenario is more likely to occur in a higher bmi patient population, which this patient was.No additional medical intervention, beyond the application of the topical cream, was recommended by physician.The ifu instructs the user to "during ablation, the saline temperature continues to rise until 90°c is reached, and the user must continuously: monitor the cervical seal for fluid loss and to confirm a tight cervical seal.Maintain a stable procedure sheath position throughout the procedure.Monitor the screen for fluid loss level." warnings and cautions related to an event of this nature are included in the ifu for genesys hta, including but not limited to: "the physician must maintain control of the procedure sheath (i.E., not hand off to another individual) for the duration of the treatment to avoid compromising the cervical seal.A compromise of the cervical seal could result in fluid leakage through the cervix, which could result in thermal injury to surrounding tissue"."do not place the procedure sheath tubing over the patient's leg or in contact with any part of the user's or patient's anatomy, as the tubing carries hot fluid and contact could result in thermal injury.The temperature of the tubing could be up to 55°c." "throughout the procedure, carefully observe the junction of the procedure sheath with the external cervical os to confirm a tight cervical seal and that there is no fluid leakage".The procedure set was discarded and was not returned for evaluation.The lot number of the procedure set was not provided; therefore, a device history record review could not be performed.However, all surgical procedure sets meet all qa specifications when released, including adequate sterilization.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|