| Model Number |
391235 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
After the first orsiro drug-eluting stent system could not pass the moderately calcified lesion (85 percent stenosis degree) in the moderately tortuous mid lad (reported separately) the second orsiro was selected for treatment.Again the the lesion could not be crossed with the device.The device was withdrawn.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned instrument was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the stent is slightly deformed at its distal end, i.E.One strut is bent outwards.Microscopic inspection showed stent imprints on the exposed balloon surface, indicating that the bent strut was initially crimped on the balloon.Outside of the deformed zone the crimped diameter of the stent complies with the specification.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|
Search Alerts/Recalls
|