Model Number 357294
Medical Device Problem Codes Fluid/Blood Leak (1250); Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
A passeo-35 balloon catheter was selected for treatment of the popliteal artery.The device was advance over the wire to the target lesion and was inflated up to 13 bar.The balloon could not be inflated further and leaking of contrast was observed in the fluoroscopy.Blood was also observed in the syringe of the inflation device.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 69 mm.Scratches were observed in close vicinity of the tear which have likely been caused by a hard, sharpedged object such as e.G.Anatomical structure.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a pressure test and a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause for the complaint event is most likely related to the patients anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-35 5/60/80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key14364789
Report Number1028232-2022-02425
Device Sequence Number5641386
Product Code DQY
UDI-Device Identifier07640119558233
UDI-Public07640119558233
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K082933
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number357294
Device Catalogue NumberSEE MODEL NO.
Device Lot Number10210329
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-