| Model Number |
FX-410360 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Seroma (2069)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.The event of seroma is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: seroma.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "peri-implant effusion." additional event of severe scarring and inflammation was reported.The device was explanted.
|
| |
|
Search Alerts/Recalls
|