| Model Number |
SOFTECHD+22.0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has seemingly been explanted but was discarded by facility prior to returning.Furthermore, lenstec can also confirm that all production records met process controls and final inspection criteria.
|
| |
|
Event or Problem Description
|
|
Lenstec received an email from stating " original lens hd +22.0 implanted and patient was experiencing anisometropia.Removed lens and implanted hd +19.0 without need for stitches to retian th elens and no adverse event for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has seemingly been explanted but was discarded by facility prior to returning.Furthermore, lenstec can also confirm that all production records met process controls and final inspection criteria.Based on the review of the batch documentation conducted, lenstec can confirm that all procedures in the manufacturing of the device were conducted correctly, and the quality of the lens is not at fault.The lens was within the tolerance range for +22.0d.
|
| |
|
Event or Problem Description
|
|
Lenstec received an email notification stating 'original lens hd +22.0 implanted and patient was experiencing anisometropia.Removed lens and implanted hd +18.0 without need for stitches to retain the lens and no adverse event for the patient'.
|
| |
|
Search Alerts/Recalls
|