• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP INC. STEAM STER LOCKS ORANGE; INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP INC. STEAM STER LOCKS ORANGE; INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS Back to Search Results
Model Number US906
Medical Device Problem Code Unexpected Color (4055)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to aesculap inc.That a steam ster locks orange (part # us906) was used to secure a sterilization container.According to the complainant, the indicator dot failed to transition to its post sterilization color.The complaint device has been returned to the manufacturer for evaluation.No patient involvement.
 
Additional Manufacturer Narrative
Manufacturer evaluation: investigation on-going.Should relevant additional / investigation results become available, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Manufacturer evaluation: the device was returned to the manufacturer for physical evaluation.Functional testing was performed by exposing ten (10) of the locks to steam cycles; a visual examination of the chemical indicator revealed that all dots transitioned from blue to brown.Us906 is assigned a three (3) year expiration.The expiration date reflected on the bag of returned material was 30apr2022.The product was past its expiration when received for complaint investigation testing.Color transition may be less reactive as the indicator nears and passes the end of its lifetime.The color will change from blue to indicate exposure to a steam process, transitioning to a grey/blue color rather than darker brown/charcoal color in this instance.Storage conditions can also affect the final transition color.The device history records (dhr) were reviewed for all available lot numbers; the devices were found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.The reported event of indicator dot failed to transition properly was not able to be confirmed.Therefore, the most likely root cause was determined to be end user storage conditions or end user equipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEAM STER LOCKS ORANGE
Common Device Name
INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
jonathan severino
861 marcon boulevard
allentown, PA 18109
4847197287
MDR Report Key14365952
Report Number2916714-2022-00044
Device Sequence Number20247275
Product Code JOJ
UDI-Device Identifier05055049091927
UDI-Public5055049091927
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K890759
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model NumberUS906
Device Catalogue NumberUS906
Device Lot Number594722-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-