| Model Number |
1012626-39 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported stent dislodgement appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported prior to the procedure the physician wanted to test if the 10.0x39mm omnilink elite stent would advance through a 7fr sheath.The device advanced without issue; however, when pulling back resistance was met with the sheath and the stent dislodged outside the patient.A non-abbott stent was used to successfully complete the procedure.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|