| Model Number |
518-086 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A peripheral atherectomy procedure commenced to treat a slightly calcified, fibrous lesion in the patient''s mid posterior tibial (pt) artery.A spectranetics quick-cross support catheter was used to cross the lesion.It was reported that a pedal approach was used.Both the anterior tibial (at) and the pt arteries were severely tortuous.During the procedure, it was observed that the quick-cross developed a hole.The physician discontinued use of the quick-cross, additional catheters were opened, and the procedure was completed with no reported patient harm.This report captures the quick-cross in which a hole was observed during use, potential for harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient''s date of birth, age unk.The device has been returned to the manufacturer, but the evaluation has not yet begun.A supplemental mdr will be submitted when device evaluation and investigation are completed.
|
| |
|
Additional Manufacturer Narrative
|
|
B5): this event is no longer reportable after device evaluation and investigation completed.G3): the device evaluation and investigation were completed on 12 may 2022.H3): the device was returned and evaluated by a cross functional team.Biologics were seen throughout the quick-cross device.The quick-cross was wavy in appearance 3.5 inches from its distal tip, as if the device was pushed up against hard material causing the quick-cross''s material to wave.Just proximal to the tip, there was a slight bulge in the quick-cross, but no significant damage was observed.All marker bands were present.Using water, the inner lumen was flushed.Water was seen leaking out of the inner lumen distal to the proximal hub.The quick-cross was twisted and kinked with multiple holes observed; this damage extended 2 cm from the proximal hub.H6): the damage likely occurred due to twisting motion and excessive force, causing the twists, kinks and holes.This has been determined to be a use related failure.
|
| |
|
Event or Problem Description
|
|
This event is no longer reportable for hole in catheter - potential for harm.Per the returned device analysis, the damaged areas of the quick-cross were found to be on the portion of the device near the proximal hub (not within the working length).
|
| |
|
Additional Manufacturer Narrative
|
|
This case was reviewed and investigated according to the manufacture¿s policy.Although the supplemental report #1 stated the event is no longer reportable, product information was found to be incorrect.Upon further review, it was determined that the information listed in the initial mdr were incorrect.Block d1) the brand name was updated to spectranetics quick-cross extreme support catheter (from spectranetics quick-cross support catheter).Block d4) the model number was updated to 518-086 (from 518-065).The udi was updated to (b)(4) (from (b)(4).Block g4) 510(k) number updated to k150201 (from k072750).Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|