• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SFX SYM PDS+ UNI VIO 18IN 0 S/A OS-6; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SFX SYM PDS+ UNI VIO 18IN 0 S/A OS-6; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Model Number SXPP1A200
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint number: (b)(4).Date sent to the fda: 5/11/2022 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The following information was requested, but was unavailable: please provide procedure name and date no further information is available.Please provide lot number =>no further information is available.Please provide the return status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details.We regularly contact with sales rep about the device returning.No further information will be provided.Investigation summary: the product was returned for evaluation.Ethicon inc conducted a visual inspection of the sample returned.Visual analysis of the returned product revealed that one opened sample product code sxpp1a200 was received for evaluation.Upon visual inspection of the returned sample, the suture barbs were to be damaged at the tips, in addition, the fixation tab was observed to have a side broken.After further evaluation of the fixation tab crimping marks were observed on the tab possibly from a surgical device.It should be noted that a 100% inspection is performed via automotive vision system to ensure the tab is present and complete during manufacturing.The manufacturing records couldn't be reviewed as the batch number is unknown.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name -irgacare.Active ingredient(s)- triclosan.Dosage form - suture/solid/parenteral.Strength - = 2360 g /m.
 
Event or Problem Description
It was reported that a patient underwent a an unknown surgery on an unknown date and barbed suture was used.It was noticed that there was only a half of the fixation tab before use, so it was not used for the surgery.There were no patient consequences reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX SYM PDS+ UNI VIO 18IN 0 S/A OS-6
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14366290
Report Number2210968-2022-03629
Device Sequence Number2046596
Product Code NEW
UDI-Device Identifier10705031227668
UDI-Public10705031227668
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K113004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSXPP1A200
Device Catalogue NumberSXPP1A200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Initial Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-