• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG., SAFELIGHT FIBER OPTIC CABLE CLEAR, 5MM X 10FT 3.05M; LIGHT SOURCE, FIBEROPTIC, ROUTINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG., SAFELIGHT FIBER OPTIC CABLE CLEAR, 5MM X 10FT 3.05M; LIGHT SOURCE, FIBEROPTIC, ROUTINE Back to Search Results
Model Number 0233050100
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was loss of light.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
The device manufacturer date is not known.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: loss of light probable root cause: intermittent electrical component contact in safelight circuit, reed switch assembly malfunction use error the reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event or Problem Description
It was reported that there was loss of light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., SAFELIGHT FIBER OPTIC CABLE CLEAR, 5MM X 10FT 3.05M
Common Device Name
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14366346
Report Number0002936485-2022-00234
Device Sequence Number4974936
Product Code FCW
UDI-Device Identifier07613327051650
UDI-Public07613327051650
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0233050100
Device Catalogue Number0233050100
Device Lot Number164807
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/15/2022
Supplement Date Received by Manufacturer04/15/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-