Model Number CPHV
Medical Device Problem Code Perivalvular Leak (1457)
Health Effect - Clinical Codes Endocarditis (1834); Hypoxia (1918); Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 03/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer was informed of this event through the patient tracking department.Based on the information received on the patient's implant forms, a carbomedics top hat valve (s5-023) was implanted on (b)(6) 2022.The device was explanted on (b)(6) 2022 and replaced with another top hat s5-023.No allegation of a device malfunction nor serious injury was received from the hospital regarding this event.Per the operational report provided, the device was explanted due to endocarditis.The patient had positive blood cultures with coag-negative staphylococcus.Patient was previously noted with an echodensity on the atrial septum.Recent transthoracic echocardiogram showed the gradient increase across both mitral and aortic valves.Aortic insufficiency was also noted.Since the patient had a recent severe deterioration of respiratory function, it was decided to to attempt the surgery to recognize if the patient's sepsis can be better controlled and to rule out valvular dysfunction due to infection.During the tee performed in or, both mitral and aortic valves seemed to be opening and closing normally.While there was increased gradient across the valve, no severe dysfunction was noted.A small paravalvular leak across the aortic valve and a thickening of atrial septum (as noted before) was noticed.No other vegetations were seen on the valve.Based on the medical note, the perivalvular leak was believed to be due to endocarditis.The patient seemed to be deteriorating despite the good antibiotic coverage.Thus, the medical decision was made to continue with the operation.During the operation, once the previous sternotomy incision was incised, there was some edema of the tissues with no obvious infection of the soft tissues or sternum.Patient was heparinized and other steps of the operation were performed.As reported during the aortic valve inspection, no vegetations or visible infection were seen on the aortic valve or annulus.Leaflets were opening and closing normally.A tiny gap between the annulus and sewing ring on the noncoronary side was noticed as a small nerve hook could pass through.The aortic valve was sharply incised and the tissue from the valve and annulus was sent for culture, but no infection on the annulus itself was visible during the surgery.The annulus was carefully debrided and all the pledgets were removed form the previous surgery, sizing was performed and size 23 carbomedics was chosen one more time to be implanted.A small strip of teflon was left between the valve ring and the aortic annulus and on the noncoronary side to prevent from any paravalvular leak in this area.A mitral valve replacement and a root enlargement were performed during the same procedure.The patient was taken to icu in stable condition.
 
Additional Manufacturer Narrative
Unknown device disposition.
 
Additional Manufacturer Narrative
A complete manufacturing and material records review for the device has been performed.The results confirmed that the device satisfied all material, visual and performance standards required at the time of manufacture and release.Since the device was not returned (unknown disposition), no further investigation can be performed at this time.Based on the information available, this carbomedics top hat valve (s5-023) was explanted after one month due to endocarditis.As reported, during the re-do procedure, the aortic valves seemed to be opening and closing normally.And no severe dysfunction was noted.Furthermore, the document review performed (that includes a review of the sterilization of the device) did not highlight any manufacturing deficiencies.Thus, based on the information available and investigation performed, there is no element suggesting a direct device relationship between the event and the device.Patient-related factors, or a nosocomial infection, could be the most probable root cause for the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBOMEDICS TOP HAT MECHANICAL HEART VALVE
Common Device Name
MECHANICAL HEART VALVE PROSTHESIS
Manufacturer (Section D)
CORCYM S.R.L.
strada crescentino
saluggia, vercelli
Manufacturer (Section G)
CORCYM S.R.L.
strada crescentino
saluggia, vercelli
Manufacturer Contact
francesca crovato
5005 north fraser way
burnaby, bc 
MDR Report Key14366634
Report Number3005687633-2022-00122
Device Sequence Number13949009
Product Code LWQ
UDI-Device Identifier08022057012913
UDI-Public(01)08022057012913(240)S5-023(17)261005
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P900060
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/05/2026
Device Model NumberCPHV
Device Catalogue NumberS5-023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer05/13/2022
Initial Report FDA Received Date05/11/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
Patient SexMale
-
-