• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", BIPOLAR, 5 X 330 MM, PREPARATION FORCEPS, MARYLAND, SHORT; HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", BIPOLAR, 5 X 330 MM, PREPARATION FORCEPS, MARYLAND, SHORT; HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR Back to Search Results
Model Number WA64320C
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 04/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic myoma enucleation procedure the jaws insert did not work in bipolar mode when combined with the erbe vio3 electrosurgical unit.After exchange of the vio3 electrosurgical unit with an erbe vio 200d device, the procedure could be completed but the patient suffered increased blood loss and the surgery time was prolonged.
 
Additional Manufacturer Narrative
The suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation found distinct signs of wear and tear on the device as well as traces of a third-party repair attempt.However, no malfunction of the device was determined and it was found to be working correctly.Based on the information available, the exact cause of the reported phenomenon and the patient¿s outcome could not be determined and is being judged as unknown.A manufacturing and quality control review was performed for the affected lot number of the jaws insert without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWS "HIQ+", BIPOLAR, 5 X 330 MM, PREPARATION FORCEPS, MARYLAND, SHORT
Common Device Name
HAND INSTRUMENTS, STANDARD, BIPOLAR AND MONOPOLAR
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key14369374
Report Number9610773-2022-00168
Device Sequence Number2006647
Product Code GEI
UDI-Device Identifier04042761062442
UDI-Public04042761062442
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K984417
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA64320C
Device Catalogue NumberWA64320C
Device Lot Number115W
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ERBE VIO 200D ELECTROSURGICAL UNIT.; ERBE VIO3 ELECTROSURGICAL UNIT.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-