• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC BOLT FOR FEMORAL NECK SYS 100 LENGTH-S; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC BOLT FOR FEMORAL NECK SYS 100 LENGTH-S; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 04.168.300S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional narrative: initial reporter is a j&j employee.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2022, the patient underwent removal of a femoral neck system and conversion to total hip arthroplasty due to the implant was subsequently beginning to cut out.The patient was previously treated for a femoral neck fracture with the femoral neck system on (b)(6) 2020.The anti-rotation screw and bolt of the fns system were observed to be cutting out superiorly as the femoral head collapsed into varus.This report is for one bolt for femoral neck sys 100 length-s.This is report 2 of 2 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOLT FOR FEMORAL NECK SYS 100 LENGTH-S
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen 2540
SZ   2540
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14371215
Report Number2939274-2022-01733
Device Sequence Number11258217
Product Code KTT
UDI-Device Identifier10886982274762
UDI-Public(01)10886982274762
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K172872
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.168.300S
Device Catalogue Number04.168.300S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/13/2022
Initial Report FDA Received Date05/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.0 TI LCKNG SCR SLF-TPNG T25 SD 40-SILE; ANTIROTATION SCR FEM NECK 100 LNTH - S; FEMORAL NECK SYSTEM PL 1 HOLE - STERILE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
Patient SexUnknown
-
-