• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 18 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 18 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number G34323
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter name and address: (b)(6).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, an advance 18 lp low profile balloon catheter would not hold inflation.The balloon was negatively prepped prior to insertion but not inflated.Upon inflation under fluoroscopy, using another manufacturer's inflation device, the balloon would not reach the intended pressure, but partial inflation was noted.The device was removed and tested outside the patient's body, showing leakage and a drop in pressure due to a pinhole.A new balloon was used to complete the procedure.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures.The patient did not experience any adverse effects due to the occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Summary of event: as reported, an advance 18 lp low profile balloon catheter would not hold inflation.The balloon was negatively prepped prior to insertion but not inflated.Upon inflation under fluoroscopy, using another manufacturer's inflation device, the balloon would not reach the intended pressure, but partial inflation was noted.The device was removed and tested outside the patient's body, showing leakage and a drop in pressure due to a pinhole.A new balloon was used to complete the procedure.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures.The patient did not experience any adverse effects due to the occurrence.Investigation evaluation: reviews of the complaint history, device history record, drawings, instructions for use (ifu), quality control procedures, and specifications were conducted during the investigation.A visual inspection and functional test of the returned device were also conducted.The complaint device was returned to cook for investigation.The balloon was leaking near the proximal bond of the balloon.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the device instructions for use (ifu), which warn, ¿if balloon pressure is lost and/or balloon rupture occurs, deflate balloon and remove balloon and sheath as a unit.¿ based upon the available information and results of the investigation, cook has concluded that a component failure unrelated to manufacturing contributed to this incident.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 18 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14371982
Report Number1820334-2022-00713
Device Sequence Number2006752
Product Code LIT
UDI-Device Identifier10827002343235
UDI-Public(01)10827002343235(17)240913(10)14207650
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K130293
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/13/2024
Device Model NumberG34323
Device Catalogue NumberPTA4-18-150-4-2
Device Lot Number14207650
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer07/22/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
B-BRAUN SWITCHBACK INFLATION DEVICE
Patient SexUnknown
-
-