EXACTECH, INC. NOVATION GXL LNR +5 LAT 40MM ID GROUP 4 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
NOVATION GXL LNR +5 LAT 40MM ID GROUP 4 CUPS |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
| Date of Event |
04/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitants medical products: element-stem, collared w/ha, std offset, sz 11 (cat# 164-03-11), nv crown cup clstr hole 60mm group 4 (cat# 180-01-60), cocr femoral head 40mm, +0 offset 12/14 (cat# 142-40-00), 3.2mm drill bit 20mm 1pk (cat# 101-05-20), alteon 6.5mm screw, 15mm (cat# 180-65-15).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, this (b)(6) patient presented to the emergency room with severe right hip pain.Ct and x-ray imaging show evidence of significant acetabular and femoral osteolysis and eccentricity of the femoral head in the acetabular liner.X-ray received.No additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, the patient involved meets the following risk criteria for early prosthesis wear/osteolysis, implanted with a lateralized liner, and the combination of largest available femoral head and/or thinnest available acetabular liner was used.The most likely cause for the prosthesis wear/osteolysis reported is a combination of the risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear.
|
| |
|