• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC SAFE-T PLUS ATRAUMATIC ADULT LP TRAY; LUMBAR PUNCTURE TRAY (ADULT & PEDIATRIC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC SAFE-T PLUS ATRAUMATIC ADULT LP TRAY; LUMBAR PUNCTURE TRAY (ADULT & PEDIATRIC) Back to Search Results
Model Number 4306CWSP
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
 
Event or Problem Description
It was reported by customer product has issue.Email verbatim: we have another lumbar puncture tray cat# 4306cwsp that we had an issue with that was related to the recent recall.Lot number 0001424764.If you could please answer the following questions, so that i can open up a new complaint for this.Can you please provide additional details in regards to the issue? manometer would not hold pressure and was leaking at connection between tubing and manometer.Can you please provide the lot number? 0001424764.Can you please provide the material number? cat# 4306cwsp.Can you please provide the date of event? (b)(6) 2022.Would you happen to have photos you can share? no photos.Would you happen to have samples to return? if so, please provide a direct address and phone.We only have one case with the same lot number unopened.What was the impact to the patient? could not get a pressure reading for the provider.How was the issue resolved? patient will need to come back for another procedure, if the provider wants to proceed again.Was there a need to repeat the procedure? unknown at this time.Is there a photo or sample available showing the reported issue for the evaluation? no.Was they able to complete the diagnosis correctly? unknown at this time.
 
Additional Manufacturer Narrative
(b)(4) follow-up emdr for device evaluation: one sample was provided to our quality team for investigation.A visual inspection of the manometer lower opening confirmed the irregular surface of the manometer bottom related to the failure mode of leakage therefore, the reported failure mode was confirmed.A review of the internal manufacturing device records and raw material history files for the reported lot number 0001424764 was performed and no recorded quality problems or rejections related to this incident were found.Based on the quality team's investigation, a growing trend in reports of this failure mode was identified.A corrective action project was opened to further investigate these defects.A voluntary recall pi-21-4292-fa was issue for the leak failure mode in the lumbar puncture trays.Please see attached recall notification letter.Complaints received for this device and defect will continue to be monitored by our quality team for signs of emerging trends.H3 other text: see manufacture narration.
 
Event or Problem Description
It was reported by customer product has issue.Email verbatim: we have another lumbar puncture tray cat #4306cwsp that we had an issue with that was related to the recent recall.Lot number 0001424764.If you could please answer the following questions, so that i can open up a new complaint for this.Can you please provide additional details in regards to the issue? manometer would not hold pressure and was leaking at connection between tubing and manometer can you please provide the lot number? 0001424764.Can you please provide the material number? cat #4306cwsp.Can you please provide the date of event? april 25, 2022.Would you happen to have photos you can share? no photos.Would you happen to have samples to return? if so, please provide a direct address and phone.We only have one case with the same lot number unopened.What was the impact to the patient? could not get a pressure reading for the provider.How was the issue resolved? patient will need to come back for another procedure, if the provider wants to proceed again.Was there a need to repeat the procedure? unknown at this time.Is there a photo or sample available showing the reported issue for the evaluation? no.Was they able to complete the diagnosis correctly? unknown at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE-T PLUS ATRAUMATIC ADULT LP TRAY
Common Device Name
LUMBAR PUNCTURE TRAY (ADULT & PEDIATRIC)
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
anna wehrheim
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key14373051
Report Number1625685-2022-00050
Device Sequence Number16655600
Product Code PWY
UDI-Device Identifier10885403229008
UDI-Public(01)10885403229008
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model Number4306CWSP
Device Catalogue Number4306CWSP
Device Lot Number0001424764
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-