• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 36MM I.D. SIZE E HIGH WALL LINER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 36MM I.D. SIZE E HIGH WALL LINER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a hip procedure, the surgeon has trouble getting the liner to insert into the cup.After several hits, the surgery was finally able to get the liner in and completed the surgery as intended.There was no reported harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant products: 110017103 ¿ g7 finned shell ¿ 7144909.Report source (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for the investigation as the product was used and remains implanted.The investigation is in process.Once the investigation is complete, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: view of the right hip demonstrates a right total hip arthroplasty with surgical skin staples suggesting recent placement.Cement fixation of the acetabular cup.Mild subcutaneous emphysema.No radiolucency or fracture.Review of the device history record(s) identified no deviations or anomalies during manufacturing.No device problem was found.As the liner was successfully assembled and implanted with no patient harm or complication.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36MM I.D. SIZE E HIGH WALL LINER
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14374993
Report Number0001822565-2022-01413
Device Sequence Number20330975
Product Code LPH
UDI-Device Identifier00889024520486
UDI-Public(01)00889024520486(17)261209(10)65257223
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K190660
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number30123605
Device Lot Number65257223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Patient Age63 YR
Patient SexMale
-
-