• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE); PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE); PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBP-125MICRO145
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 04/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The crown of the diamondback 360 peripheral orbital atherectomy device (oad) became entrapped in a calcified area of the patient's peroneal artery, and the oad stalled.The vessel was profusely diseased, approximately 2.5 mm in diameter, and was 80% to 90% stenosed.The physician removed the viperwire advance guide wire and detached the saline line from the device.The saline sheath of the driveshaft was removed.The crown remained in vivo.Surgery was performed to access the peroneal artery.The stuck segment of the oad was removed from the patient.The patient was stable.
 
Additional Manufacturer Narrative
The diamondback 360 peripheral orbital atherectomy device (oad) and viperwire advance guide wire were returned to csi for analysis.The distal segments of the driveshaft and guide wire were not returned.Analysis of the returned portions of the device did not identify any damage or abnormalities that would have contributed to the reported complaint.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE)
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
kim mcclure
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key14377985
Report Number3004742232-2022-00117
Device Sequence Number19334933
Product Code MCW
UDI-Device Identifier10850000491172
UDI-Public(01)10850000491172(17)220731(10)337328
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model NumberDBP-125MICRO145
Device Catalogue Number7-10057-01
Device Lot Number337328
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2022
Initial Report FDA Received Date05/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexFemale
Patient Weight91 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-