Model Number R SERIES |
Device Problems
No Display/Image (1183); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device powerd up in the incorrect mode and the display blanked out.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to an integrated circuit on the digital board.The digital board was replaced to resolve the customer's report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|