• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. PROFEMUR® NECK A/R 8DG LONG COBALT CHROME; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. PROFEMUR® NECK A/R 8DG LONG COBALT CHROME; HIP COMPONENT Back to Search Results
Model Number PHAC1234
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Codes Neuropathy (1983); Pain (1994); Metal Related Pathology (4530)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly pt.Now has elevated metal ions.Originally part of ide / ceramic on metal articulation system clinical study but not any longer.Additional information received on (b)(6) 2022: patient developed adverse symptoms including, but not limited to, progressively worsening pain in and around his right hip joint, as well as neurologic symptoms including peripheral neuropathy, tinnitus, some balance/coordination issues, a heal fracture as a direct result of those balance/coordination issues, and b 12 deficiency.Laboratory test results reported on (b)(6) 2020, revealed elevated serum chromium and blood cobalt levels in patient.On (b)(6) 2021, patient had a revision surgery performed on his right hip by (b)(6) m.D., at (b)(6) hospital, (b)(6).Early in his (b)(6) 2021, operative report, dr.Jones noted, "metallosis debris at the external rotators." after removal of the conserve® femoral head from the profemur® plus modular neck, dr.(b)(6) noted "a fair amount of black corrosion discoloration at the taper on both components." in the (b)(6) 2021, revision surgery, dr.(b)(6), removed the wright medical profemur® hip system, including the profemur® plus cocr modular neck and profemur® plasma z stem, as well as the conserve® femoral head, that had been implanted in plaintiff's right hip in 2010, were removed, and replaced with a femoral stem, femoral head, and a dual mobility bearing, all devices from zimmer biomet.In the same revision surgery, the instrument used to extract the stem had a failure and it couldn't' be used.Therefore, an extended trochanteric osteotomy was necessary.This problem has been captured and investigated under microport orthopedics incident number: (b)(4).
 
Event or Problem Description
Allegedly pt.Now has elevated metal ions.Originally part of ide / ceramic on metal articulation system clinical study but not any longer.Additional information received on (b)(6) 2022: patient developed adverse symptoms including, but not limited to, progressively worsening pain in and around his right hip joint, as well as neurologic symptoms including peripheral neuropathy, tinnitus, some balance/coordination issues, a heal fracture as a direct result of those balance/coordination issues, and b 12 deficiency.Laboratory test results reported on (b)(6) 2020, revealed elevated serum chromium and blood cobalt levels in patient.On (b)(6) 2021, patient had a revision surgery performed on his right hip by daniel a.Jones, m.D., at cleveland clinic martin south hospital, stuart, florida.Early in his (b)(6) 2021, operative report, dr.Jones noted, "metallosis debris at the external rotators." after removal of the conserve® femoral head from the profemur® plus modular neck, dr.Jones noted "a fair amount of black corrosion discoloration at the taper on both components." in the (b)(6) 2021, revision surgery, dr.Jones, removed the wright medical profemur® hip system, including the profemur® plus cocr modular neck and profemur® plasma z stem, as well as the conserve® femoral head, that had been implanted in plaintiff's right hip in 2010, were removed, and replaced with a femoral stem, femoral head, and a dual mobility bearing, all devices from zimmer biomet.In the same revision surgery, the instrument used to extract the stem had a failure and it couldn?t be used.Therefore, an extended trochanteric osteotomy was necessary.This problem has been captured and investigated under microport orthopedics incident number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® NECK A/R 8DG LONG COBALT CHROME
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key14378451
Report Number3010536692-2022-00194
Device Sequence Number17391085
Product Code JDL
UDI-Device IdentifierM684PHAC12341
UDI-PublicM684PHAC12341
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K091423
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPHAC1234
Device Catalogue NumberPHAC1234
Device Lot Number0201053867
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/28/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-