U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH*PROBE PLUS II SHAFT; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH*PROBE PLUS II SHAFT; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number EPS07
Medical Device Problem Code Arcing of Electrodes (2289)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.
 
Event or Problem Description
It was reported that, during a laparoscopic pneumonectomy, the device could not energize properly and did not function.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Additional Manufacturer Narrative
(b)(4) date sent: 6/20/2022 d4 batch #: v95v39 investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the eps07 device was received with the electrode tip bent and the shaft sheath damaged at the instrument tip.A bent electrode could have caused this damage to the sheath.Caution should be taken not to retract the electrode into the sheath when it is bent.Due to this condition we were unable to investigate further the reported cautery isssues.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: do not attempt to bend, sharpen, or otherwise alter the shape of the electrode tip.Doing so may cause electrode tip failure and user or patient injury.Please reference the instruction for use for more information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
MDR Report Key14378855
Report Number3005075853-2022-03128
Device Sequence Number4990455
Product Code GCJ
UDI-Device Identifier10705036012511
UDI-Public10705036012511
Combination Product (Y/N)N
PMA/510(K) Number
K912492
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2026
Device Model NumberEPS07
Device Catalogue NumberEPS07
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date06/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-