Model Number APB-2-2-3D-ES
Medical Device Problem Codes Detachment of Device or Device Component (2907); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that the coil got stuck inside the microcatheter, eventually a detachment did occur inside the catheter.The device was prepared as indicated in the package insert.Ancillary devices include: headway 17 microcatheter.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3:¿ as found condition: the axium prime coil was returned for analysis within a shipping box, within a primary plastic resealable pouch, within an opened axium prime outer carton (221070140), and within a secondary plastic resealable pouch.A (microvention terumo) headway 17 catheter was returned with the axium prime coil.¿ visual inspection/damage location details: the axium prime pusher¿s actuator interface (ai) and hypotube break indicator (hbi) were found intact.No bends or kinks were found with the pusher.The release wire coin was found against the lumen stop, the shield coil was in good condition, and the implant coil was detached.The implant coil was not returned.The release wire distal tip was extending out from the distal end of the pusher.The release wire was removed from the pusher and the coin was found in good condition.The pusher lumen stop and retainer ring inner diameters were examined and found in good condition.The headway catheter was examined.The catheter hub was found in good condition.No kinks were found with the catheter body.The catheter distal tip was found bent.¿ testing/analysis: the headway catheter was flushed, blood and water exited from the distal tip.An in-house 0.016¿ mandrel was inserted through the headway catheter without issue.¿ conclusion: based on the device analysis and reported information, the customer¿s ¿premature detachment¿ report was confirmed.As the axium prime coil was advanced against resistance this would have likely contributed to the event.Advancing the device against resistance cause the implant coil to detach.The implant coil was not returned; therefore, analysis could not be performed and conformance to specification could not be assessed.Possible contributing factors to ¿coil resistance¿ include using an incompatible device for delivery, tortuous anatomy, failure to maintain a continuous flush or failure to hydrate the axium prime coil before use.However, additional information was not provided.The headway 17 catheter has a labeled id (inner diameter) of 0.017¿, as per an online source.Per the axium prime coil ifu (instructions for use): ¿axium prime detachable coils should be delivered through a microcatheter with a minimum inside diameter of 0.0165¿-0.017¿ with two marker bands¿.Therefore, the headway 17 catheter was found to be compatible for use with the axium prime coil.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS D/B/A EV3 NEUROVASCULAR
9779 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS D/B/A EV3 NEUROVASCULAR
9779 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key14379857
Report Number9612164-2022-01790
Device Sequence Number12448282
Product Code KRD
UDI-Device Identifier00847536021174
UDI-Public00847536021174
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/30/2023
Device Model NumberAPB-2-2-3D-ES
Device Catalogue NumberAPB-2-2-3D-ES
Device Lot Number221070140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-