• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. PROFEMUR® TL STEM SZ 5; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. PROFEMUR® TL STEM SZ 5; HIP COMPONENT Back to Search Results
Model Number PRTL0025
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Metal Related Pathology (4530)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly, the patient has been suffering moderate pain, and the implants were revised for metallosis, according to the surgeon.The modular neck trunion (fitting in to the stem pocket) was noticeably "blackened".All implants were removed, and replaced with a thr cement spacer.Unfortunately it wasn't possible to remove the liner from the procotyl l shell, so i cant provide the lot number for the biolox liner.
 
Event or Problem Description
Allegedly, the patient has been suffering moderate pain, and the implants were revised for metallosis, according to the surgeon.The modular neck trunion (fitting in to the stem pocket) was noticeably "blackened".All implants were removed, and replaced with a thr cement spacer.Unfortunately it wasn't possible to remove the liner from the procotyl l shell, so i cant provide the lot number for the biolox liner.Additional information received on 05/13/2022 from njr: njr advised that the patient was reviewed due to infection and adverse soft tissue reaction to particulate debris.Revision njr number: (b)(4).Site: r.Primary asa: p2 - mild disease not incapacitating.
 
Additional Manufacturer Narrative
Section b.5:additional information in description / section f.6: awareness date update / section h.6 health code (1930) added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® TL STEM SZ 5
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key14380060
Report Number3010536692-2022-00195
Device Sequence Number6725198
Product Code JDI
UDI-Device IdentifierM684PRTL00251
UDI-PublicM684PRTL00251
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K060358
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberPRTL0025
Device Catalogue NumberPRTL0025
Device Lot Number1746041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/13/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-