Model Number 434897
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Arrest (1762)
Date of Event 04/21/2022
Type of Reportable Event Death
Event or Problem Description
The pk papyrus covered stent system was selected for treatment of a type iii perforation in the lad.After deployment of the covered stent size 4.5/20 loss flow in the dominant left cx occurred, which required angioplasty to restore the flow.A second perforation was noticed in the mid lad distally, which was covered with a pk papyrus stent size 2.5/20.At the edge of this pk papyrus stent there was a distal dissection which was treated with a des stent.(this event is reported separately.) despite all efforts the patient continued to deteriorate.Patient had multiple comorbidities and the likelihood of meaningful survival was low to none.Patient passed away shortly after the intervention.
 
Additional Manufacturer Narrative
The affected pk papyrus stent system was not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report, procedure note, discharge summary) was reviewed to establish whether a device deficiency contributed to this event.Review of the product release documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause was identified.The treating physician rated the outcome as both device- and procedure-related complication.Please note that the ifu lists the risk of treatment-related occlusion of vital coronary artery side branches as contraindication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 4.5/20
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key14380220
Report Number1028232-2022-02469
Device Sequence Number5150134
Product Code NIV
UDI-Device Identifier07640130447257
UDI-Public07640130447257
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number434897
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03211126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age81 YR
Patient SexFemale
-
-