Catalog Number 0116311
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Hernia (2240)
Date of Event 04/21/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported, post-implant of the 3dmax mid, the patient experienced hernia recurrence and underwent additional surgery for hernia repair.Based on the information available, no conclusions can be made.Hernia recurrence is a known inherent risk/use of the device, listed as a possible complication in the adverse reactions section of the instructions for the use (ifu) as a possible complication.Review of the manufacturing records confirms product was manufactured to specification.To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in november, 2020.Not returned - remains implanted.
 
Event or Problem Description
As reported, during a robotic assisted left inguinal herniorrhaphy procedure on (b)(6)2021, the patient was implanted with bard/davol 3dmax mid lg, left.During this procedure the patient also underwent a stitch umbilical herniorrhaphy with carter thomason needle passer, 0 prolene.As reported, the patient underwent a robotic assisted repair of recurrent left inguinal hernia repair procedure on (b)(6) 2021 during which a 3dmax mid mesh was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DMAX MID
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14380469
Report Number1213643-2022-00224
Device Sequence Number11980536
Product Code FTL
UDI-Device Identifier00801741200755
UDI-Public(01)00801741200755
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K200818
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/28/2022
Device Catalogue Number0116311
Device Lot NumberHUEXAB02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2022
Initial Report FDA Received Date05/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;