| Model Number |
SCF-295 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" against left device.Additional information notes, calcification and folds the device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation:.Analysis of the returned device identified: weight to spec, deformation device, white particles in the shell and creases fold.Cloudy and voids in the gel observed after autoclave cycle the unit is not rupture based on the device analysis the final assessment is: no issues found related with the manufacturing process.Additional, changed, and/or corrected data: d9, g2, h3, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" against left device.Additional information notes, calcification and folds the device has been explanted.
|
| |
|
Search Alerts/Recalls
|