| Model Number |
MICL 13.7 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Reportedly a 13.7mm micl13.7 implantable collamer lens of -3.5 diopter was removed due to high vault.It was replaced with a shorter length lens on (b)(6) 2022.The lens was removed on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient identifier: unknown,age or date of birth:unknown, sex::unknown, weight: unknown, ethnicity:unknown, race:unknown.Date of event: unk.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the reporter indicated that a 13.7mm micl13.7 implantable collamer lens of -3.5 diopter was implanted into the patient's left eye (os) on (b)(6) 2022.Reportedly excessive vault, elevated iop, and narrow angles were found.The lens was explanted on (b)(6) 2022, and replaced on the same day in a separate surgery with a shorter length lens of a different model.This resolved the problem.H6- health effect- clinical code 4581: narrow angles.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|