| Model Number |
PSCST30 |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Unknown, as information was asked but not provided.Date implanted: not applicable, as the cartridge is not an implantable device.Date explanted: not applicable, as the cartridge is not an implantable device.It was indicated that both the pscst cartridge and the z9002 iol are not being returned for evaluation as they were both discarded therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.An attempt was made to obtain missing information.However; the account did not provide some information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the unloader did something to damage the intraocular lens (iol).Through follow-up, additional information was received clarifying the tip of the pscst unfolder (cartridge) was observed under microscope by surgeon to be damaged and have a jagged edge.The z9002 lens was already in the inserter, so it could not be reused with a new inserter.The damaged tip was not visible to the naked eye, only under the microscope at the time of insertion.Cartridge tip had patient contact.It was confirmed the pscst cartridge was inspected inspected under microscope prior to use.A new lens with the same model and diopter size and inserter were used to complete the procedure.There was no unplanned incision enlargement, no serious patient injury, no unplanned suture(s), and no unplanned vitrectomy.Patient outcome post-procedure was reported as positive.The iol and the cartridge are not available for return as they were both discarded.No further information is available.
|
| |
|
Search Alerts/Recalls
|
|
|