• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 18 MG BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 18 MG BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 368589
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 18 mg blood collection tube there was under-fill or low draw of a tube with blood.This event occurred 10 times.The following information was provided by the initial reporter.The customer stated: "the customer is experiencing a low draw volume.".
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H.6.Investigation summary: bd received 104 samples from the customer in support of this complaint.(b)(4) sample tubes were draw tested and the issue of underfill was not observed as all tubes were within specification limits.Product has not expired at the time of testing.Additionally, (b)(4) retention tubes samples from the bd inventory were functionally tested and the issue of underfill was not observed as all tubes were within specification limits.Bd was unable to confirm and/or duplicate the customer¿s indicated failure mode because the defect was not observed in the sample and retention testing.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 18 mg blood collection tube there was under-fill or low draw of a tube with blood.This event occurred 10 times.The following information was provided by the initial reporter.The customer stated: "the customer is experiencing a low draw volume.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 18 MG BLOOD COLLECTION TUBE
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14388227
Report Number1917413-2022-00257
Device Sequence Number17358260
Product Code JKA
UDI-Device Identifier50382903685894
UDI-Public50382903685894
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
BK980011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2022
Device Model Number368589
Device Catalogue Number368589
Device Lot Number0289469
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date05/13/2022
Supplement Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-