| Model Number |
368589 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® k2 edta (k2e) 18 mg blood collection tube there was under-fill or low draw of a tube with blood.This event occurred 10 times.The following information was provided by the initial reporter.The customer stated: "the customer is experiencing a low draw volume.".
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: bd received 104 samples from the customer in support of this complaint.(b)(4) sample tubes were draw tested and the issue of underfill was not observed as all tubes were within specification limits.Product has not expired at the time of testing.Additionally, (b)(4) retention tubes samples from the bd inventory were functionally tested and the issue of underfill was not observed as all tubes were within specification limits.Bd was unable to confirm and/or duplicate the customer¿s indicated failure mode because the defect was not observed in the sample and retention testing.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® k2 edta (k2e) 18 mg blood collection tube there was under-fill or low draw of a tube with blood.This event occurred 10 times.The following information was provided by the initial reporter.The customer stated: "the customer is experiencing a low draw volume.".
|
| |
|